​ADI-SPS-870-F – Antibody Lot Comparison Data

Hi Erica,

 

We have observed a substantial lot-dependent change in the performance of antibody ADI-SPS-870-F in our established assay.

Specifically, the new lot P39BA03B25A produces substantially increased nonspecific background compared with the previous lot 01031953 in several capture configurations. Additional troubleshooting confirmed that the effect varies by capture reagent and cannot be resolved simply by diluting the antibody.

The representative experimental data are summarized below:

Capture                                  Sample    LOT 01031953   LOT P39BA03B25A  Ratio
                                                  (a.u.)         (a.u.)
------------------------------------------------------------------------------------------
Dynabeads His-Tag Isolation & Pulldown   Buffer 1        7.693            197.50   25.70
Dynabeads M-270 Epoxy, reagent 1         Buffer 1       22.370             16.63    0.74
Dynabeads M-270 Epoxy, reagent 2         Buffer 1        6.005             13.29    2.21
Dynabeads M-270 Epoxy, reagent 3         Buffer 2       27.710             87.78    3.17
Dynabeads M-270 Epoxy, reagent 4         Buffer 2       11.050             20.07    1.82

Ratio = background signal from lot P39BA03B25A divided by background signal from lot 01031953.

​

​

As discussed, we are requesting a complimentary replacement of the problematic vial with the reserved vial of lot 01031953.

Order: A62371327 / PO DR5834383
Problematic lot: P39BA03B25A
Requested replacement lot: 01031953

​

Thank you!