ADI-SPS-870-F – Antibody Lot Comparison Data
Hi Erica,
We have observed a substantial lot-dependent change in the performance of antibody ADI-SPS-870-F in our established assay.
Specifically, the new lot P39BA03B25A produces substantially increased nonspecific background compared with the previous lot 01031953 in several capture configurations. Additional troubleshooting confirmed that the effect varies by capture reagent and cannot be resolved simply by diluting the antibody.
The representative experimental data are summarized below:
Capture Sample LOT 01031953 LOT P39BA03B25A Ratio
(a.u.) (a.u.)
------------------------------------------------------------------------------------------
Dynabeads His-Tag Isolation & Pulldown Buffer 1 7.693 197.50 25.70
Dynabeads M-270 Epoxy, reagent 1 Buffer 1 22.370 16.63 0.74
Dynabeads M-270 Epoxy, reagent 2 Buffer 1 6.005 13.29 2.21
Dynabeads M-270 Epoxy, reagent 3 Buffer 2 27.710 87.78 3.17
Dynabeads M-270 Epoxy, reagent 4 Buffer 2 11.050 20.07 1.82
Ratio = background signal from lot P39BA03B25A divided by background signal from lot 01031953.
As discussed, we are requesting a complimentary replacement of the problematic vial with the reserved vial of lot 01031953.
Order: A62371327 / PO DR5834383
Problematic lot: P39BA03B25A
Requested replacement lot: 01031953
Thank you!